The Cerabar PMP71 is a digital pressure transmitter featuring a metallic membrane, ideal for various process applications including pressure, level, volume, or mass measurement in both liquids and gases. It is specifically engineered for high-pressure environments, supporting applications up to 700 bar. The device offers a Quick Setup feature, allowing for freely adjustable measuring ranges without requiring specific pressure inputs. It is designed in accordance with IEC 61508 standards, making it suitable for use in SIL2/3 safety systems, and it comes with a MID Parts Certificate, making it appropriate for custody transfer applications.
The transmitter is housed in a robust two-chamber aluminum or 316L stainless steel casing.
Available process connections include thread and flange options.
It operates within a process temperature range of -40 to +125°C (-40 to +257°F).
Measuring ranges span from -1/0 up to 700 bar (-15/0 up to +10,500 psi).
The device offers high accuracy, with a standard of ±0.05%, and a "Platinum" accuracy option of ±0.025%.
It also comes with international explosion protection certifications, WHG overfill prevention, hygienic and marine approvals, and SIL compliance.
Benefits
- Delivers exceptional accuracy, reproducibility, and long-term stability.
- Ensures maximum safety with a gas-tight feedthrough, capable of supporting SIL2/3 applications, certified to IEC 61508 standards.
- Features an easy, menu-guided setup process via a local display, supporting 4 to 20 mA with HART, PROFIBUS PA, and FOUNDATION Fieldbus communication protocols.
- Includes the HistoROM data management concept, facilitating quick and straightforward commissioning, maintenance, and diagnostics.
- The device is overload-resistant and continuously monitors functionality from the measuring cell to the electronics.
- Available with pre-mounted manifolds that are fully tested for leaks, ensuring a perfect fit.
- Offers seamless and independent system integration through HART, PROFIBUS PA, or FOUNDATION Fieldbus.
SPECIFICATIONS
Pressure Specifications
Measuring principle | Absolute and gauge pressure |
Characteristic |
Digital transmitter with piezoresistive sensor and welded metallic membrane Modular transmitter Long term stability Enhanced safety via self diagnostic functionsSecondary process barrier |
Supply voltage |
4...20 mA HART 10,5...45V DC (Non Ex): Ex ia: 10,5...30V DC PROFIBUS PA: 9...32 V DC (Non Ex) FOUNDATION Fieldbus: 9...32 V DC (Non Ex) |
Reference Accuracy |
Standard: 0.05% Platinum: up to 0.025% |
Long term stability |
0.05 % of URL/ year 0.07 % of URL/ 5 years 0.1 % of URL/ 10 years |
Process temperature |
-40°C...125°C (-40°F...257°F) |
Ambient temperature |
-50°C...85°C (-58°F...185°F) |
Measuring cell |
400 mbar...700 bar (5.8 psi...10.500 psi) relative/ absolute |
Smallest calibratable span |
5 mbar (0.075 psi) |
Vacuum resistance | 10 mbar (0.15 psi) |
Max. Turn down |
100:1 |
Max. overpressure limit | 1050 bar (15.750 psi) |
Process connection |
Thread: G1/2...G2, R1/2, MNPT1/2...MNPT2, R1/2 Flange: DN25...DN80, ASME 1"...4", JIS 10K...20K |
Material process membrane |
316L, AlloyC, Rhodium> Gold |
Material gasket | None, measuring cell welded |
Fill fluid |
Silicone oil, Inert oil |
Material housing |
Die-cast aluminum, AISI 316L |
Communication |
4...20 mA HART 1- 5V DC Low Power PROFIBUS PA FOUNDATION Fieldbus |
Certificates / Approvals | ATEX, FM, CSA, CSA C/US, IEC Ex, JPN Ex, INMETRO, NEPSI, EAC, UK Ex |
Safety approvals | SIL |
Design approvals |
EN10204-3.1 NACE MR0175 |
Marine approvals |
GL/ABS |
Drinking water approvals | NSF |
Specialities | Diagnostic functions |
Successor | PMP71B |
Continuous / Liquids Specifications
Measuring principle | Absolute and gauge pressure |
Characteristic / Application |
Digital transmitter with piezoresistive sensor and welded metallic membrane Modular transmitter Long term stability Enhanced safety via self diagnostic functions Secondary process barrier |
Specialities |
diagnostic functionalities different languages in software |
Supply / Communication |
4...20mA HART: 10,5...45V DC Ex ia: 10,5...30V DC PROFIBUS PA / FOUNDATION Fieldbus: 9...32V DC |
Accuracy | Standard: 0.05%
Platinum: up to 0.025% |
Long term stability | 0,05% of URL/year |
Ambient temperature | -50°C...85°C
(-58°F...185°F) |
Process temperature | -40°C...125°C
(-40°F...257°F) |
Process pressure / max. overpressure limit | 1050bar (15,750psi) |
Pressure measuring range | 400 mbar...700 bar
(5.8 psi...10.500 psi) |
Main wetted parts | Alloy C276
316L Rhodium |
Process connection | Threads
Flanges (DIN, ASME, ...) with flush membrane |
Max. measurement distance | 7000m (22.966ft) H2O |
Communication | 4...20 mA HART
1- 5V DC Low Power PROFIBUS PA FOUNDATION Fieldbus |
Certificates / Approvals | ATEX, FM, CSA, CSA C/US, IEC Ex, JPN Ex, INMETRO, NEPSI, EAC |
Safety approvals | SIL |
Design approvals | EN 10204-3.1
NACE MR0175, MR0103 |
Marine approval | GL/ ABS |
Drinking water approvals | NSF |
Options | HistoROM/M-Dat
4-line digital display SS- or Aluminiumhousing Separate housing |
Successor | PMP71B |
Application limits | Measuring cell: Metal
welded If pressurized, possibly use differential pressure meas-urement with two pressure transmitters (electronic dp). Observe ratio head pressure : hydrostatic pressure |
Documents