null
Endress+Hauser

Absolute and gauge pressure Cerabar PMP75 series

(No reviews yet) Write a Review
SKU:
Absolute and gauge pressure Cerabar PMP75 series
Weight:
0.00 Grams
Call us: +971 4 259 5133

Description

Application

The Cerabar PMP75 is a digital pressure transmitter equipped with a metal diaphragm seal, designed primarily for process and hygiene applications involving the measurement of pressure, level, volume, or mass in liquids or gases. It is particularly well-suited for high-pressure environments and extreme process temperatures, ranging from -70°C to +400°C (-94°F to 750°F). The device offers a Quick Setup feature with an adjustable measuring range, streamlining the commissioning process, reducing costs, and saving time. Built in accordance with IEC 61508 standards, the Cerabar PMP75 is suitable for use in SIL2/3 safety applications.

The transmitter is available with a dual-chamber housing made of either aluminum or stainless steel (316L).

Process connections: Options include threads, flanges, or hygienic connections.

Process temperature range: -70°C to +400°C (-94°F to 752°F).

Pressure range: -1/0 up to 400 bar (-15/0 up to +6,000 psi).

Accuracy: ±0.075%.

The Cerabar PMP75 comes with international certifications for explosion protection, overfill prevention (WHG), SIL compliance, hygienic approvals, and marine approvals.

 

 


 

Benefits

  • Wide selection of process connections and diaphragm materials to suit various applications.
  • The innovative TempC Membrane reduces the impact of ambient and process temperature fluctuations.
  • The HistoROM data management system enables quick and easy commissioning, maintenance, and diagnostics.
  • User-friendly menu-driven setup via local display, with options for 4 to 20mA HART, PROFIBUS PA, or FOUNDATION Fieldbus communication.
  • Enhanced safety with gas-tight feedthroughs, supporting SIL2/3 applications and certified to IEC 61508 standards.
  • Cost efficiency through a modular design that allows easy replacement of sensors, displays, or electronics.
  • The system is resistant to overload and continuously monitored from the sensor to the electronics.

SPECIFICATIONS

Pressure Specifications

Measuring principle Absolute and gauge pressure
Characteristic

Digital transmitter with piezoresistive sensor and diaphragm seal

Modular transmitter

Long term stability

Minimum oil volume process connection

Enhanced safety via self diagnostic functions

Secondary process barrier

 Supply voltage  4...20 mA HART

10,5...45V DC (Non Ex):

Ex ia: 10,5...30V DC

PROFIBUS PA:

9...32 V DC (Non Ex)

FOUNDATION Fieldbus:

9...32 V DC (Non Ex)

 Reference Accuracy  0,075% + influence of diaphragm seal
 Long term stability  0.05 % of URL/ year

0.07 % of URL/ 5 years

0.1 % of URL/ 10 years

 Process temperature  -70°C...400°C

(-94°F...752°F)

 Ambient temperature  -50°C...85°C

(-58°F...185°F)

 Measuring cell 400 mbar...400 bar

(5.8 psi...6000 psi)

relative/ absolute

 Smallest calibratable span  5 mbar (0.075 psi)
 Vacuum resistance  10 mbar (0.15 psi)
 Max. Turn down  100:1
 Max. overpressure limit  1050 bar (15.750 psi)
 Process connection  Thread:

G1/2...G2, R1/2, MNPT1/2...MNPT2, NPT1/2...NPT1

Flange:

DN25...DN100,

ASME 1"...4",

JIS 10K

Diaphragm seal

 Process connection hygienic  Tri-Clamp

DIN11851

NEUMO

Varivent

SMS

DRD

Universal adapter

 Material process membrane  316L, AlloyC,

Tantal

Rhodium> Gold

PTFE

 Material gasket  None, diaphragm welded
 Fill fluid Silicone oil,

Inert oil,

Vegetable oil,

High temperature oil,

Low temperature oil,

 Material housing  Die-cast aluminum,

AISI 316L

 Communication  4...20 mA HART

PROFIBUS PA

FOUNDATION Fieldbus

 Certificates / Approvals  ATEX, CSA C/US, IEC Ex, JPN Ex, INMETRO, NEPSI, EAC
 Safety approvals SIL 
 Design approvals  EN10204-3.1

NACE MR0103

 Hygienic approvals  3A, EHEDG
 Marine approvals  GL/ ABS
 Specialities  Diagnostic functions

TempC Membrane

 Successor  PMP71B

 Continuous / Liquids Specifications

Measuring principle Absolute and gauge pressure
Characteristic / Application

Digital transmitter with piezoresistive sensor and diaphragm seal

Modular transmitter

Long term stability

Minimum oil volume

Enhanced safety via self diagnostic functions

Secondary process barrier

Specialities

Diagnostic functionalities

Different languages in software

Supply / Communication

4...20mA HART:

10,5...45V DC

Ex ia: 10,5...30V DC

PROFIBUS PA /

FOUNDATION Fieldbus:

9...32V DC

Accuracy 0,075% + influence of diaphragm seal
Long term stability 0,05% of URL/year
Ambient temperature

-50°C...85°C

(-58°F...185°F)

Process temperature

-70°C...400°C

(-94°F...752°F)

Process pressure / max. overpressure limit

1050bar (15,200psi)
Pressure measuring range

400 mbar...400 bar

(5.8 psi...6000 psi)

Main wetted parts

Alloy C276

316L

Monel

Tantalum

PTFE-Foil

Process connection

Threads

Flanges (DIN, ASME, JIS) with flush membrane

Tri-Clamp ISO02852

Hygienic connections

Max. measurement distance 7000m (22.966ft) H2O
Communication

4...20 mA HART

PROFIBUS PA

FOUNDATION Fieldbus

Certificates / Approvals ATEX, FM, CSA, CSA C/US, IEC Ex, JPN Ex, INMETRO, NEPSI, EAC
Safety approvals

SIL

Design approvals

EN 10204-3.1

NACE MR0175, MR0103

Hygienic approvals 3A, EHEDG
Marine approval

GL/ ABS

Options

HistoROM/M-Dat

4-line digital display

SS- or Aluminiumhousing

Separate housing

Successor

PMP71B
Application limits

Measuring cell: Metal

welded

If pressurized, possibly

use differential pressure

meas-urement with two

pressure transmitters

(electronic dp). Observe

ratio head pressure :

hydrostatic pressure

 

Documents

You may want to read this product's catalog. If so, please click to view the Endress+Hauser Absolute and gauge pressure Cerabar PMP75 Series datasheet.
View AllClose