Endress+Hauser
Accurate Coriolis Flowmeter LPGmass D8EB
- SKU:
- LPGmass D8EB Series
- MPN:
- LPGmass D8EB
- Weight:
- 0.00 Grams
Description
Application
The LPGmass is specifically engineered for precise flow measurement of LPG in dispensing and truck unloading processes. It features integrated temperature measurement and intelligent conversion capabilities, enabling direct volume correction on-site. This makes the LPGmass an ideal solution for system integrators, skid builders, and equipment manufacturers.
- The measuring principle is unaffected by physical properties of the fluid, such as viscosity or density.
- Provides accurate measurements of liquefied petroleum gas for refueling and distribution purposes.
Device Properties:
- Handles flow rates up to 70,000 kg/h (2,570 lb/min).
- Offers volume flow calculations following API Table 53 standards.
- Features a robust, compact transmitter housing.
- Equipped with Modbus RS485 communication.
- Designed to meet a wide range of application requirements.
Benefits
- Superior operational safety – remains reliable even under challenging ambient conditions.
- Reduced process measurement points – offers multivariable measurement of flow, density, and temperature.
- Compact installation – requires no inlet or outlet runs, saving space.
- Space-efficient transmitter – delivers full functionality within a compact design.
- Quick setup – devices come pre-configured for fast commissioning.
- Easy maintenance – supports automatic data recovery for servicing needs.
Documents
You may want to read this product's catalog. If so, please click to view the LPGmass D8EB Series datasheet.SPECIFICATIONS
Liquids Specifications
Measuring principle | Coriolis |
Product headline | The refueling and distribution application flowmeter with easy system integration. Accurate measurement of liquefied petroleum gas in refueling and distribution applications. |
Sensor features | Excellent operational safety – reliable under extreme ambient conditions. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space-saving installation – no in/outlet run needs. Flow rates up to 70 000 kg/h (2570 lb/min). Volume flow calculation according to API table 53. |
Transmitter features | Space-saving transmitter – full functionality on the smallest footprint. Fast commissioning – pre-configured devices. Automatic recovery of data for servicing. Robust, compact transmitter housing. Modbus RS485. |
Nominal diameter range | DN 8 to 50 (³⁄₈ to 2") |
Wetted materials | Measuring tube: 1.4539 (904L) Connection: 1.4404 (316/316L) |
Measured variables | Mass flow, density, temperature, volume flow, corrected volume flow, reference density |
Max. measurement error | Mass flow (liquid): ±0.20 % Volume flow (liquid): ±0.30 % |
Measuring range | 0 to 70 000 kg/h (0 to 2570 lb/min) |
Max. process pressure | PN 40, Class 300 |
Medium temperature range | –50 to +125 °C (–58 to +257 °F) |
Ambient temperature range | –40 to +60 °C (–40 to +140 °F) |
Sensor housing material | 1.4301 (304), corrosion resistant |
Transmitter housing material | AlSi10Mg, coated |
Degree of protection | IP66/67, type 4X enclosure |
Display/Operation | No local operation Configuration via operating tools possible |
Outputs | None |
Inputs | None |
Digital communication | Modbus RS485 |
Power supply | DC 20 to 30 V |
Hazardous area approvals | ATEX, IECEx, cCSAus, NEPSI, INMETRO |
Product safety | CE, C-Tick |
Metrological approvals and certificates | Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025) |
Pressure approvals and certificates | PED, CRN |
Material certificates | 3.1‐Material |