Application

The Prosonic FMU42 Ultrasonic Level Sensor is designed for non-contact level measurement across a variety of media, including liquids, pastes, and coarse bulk materials, as well as for flow measurement in open channels or at weirs. Its versatile two-wire or four-wire configuration makes it suitable for use in storage tanks, with agitators, on stockpiles, and on conveyor belts. The sensor features an on-site display that can visualize the envelope curve for straightforward diagnostics. Additionally, it offers a linearization function (up to 32 points) to convert measured values into units of length, volume, or flow rate.

  • Measurement Type: Continuous non-contact level and flow measurement
  • Process Connection: Flange
  • Temperature Range: -40 to +80°C (-40 to +176°F)
  • Pressure Range: +0.7 to +2.5bar (+10 to +36psi)
  • Max Measuring Distance: 10m (33ft) for liquids, 5m (16ft) for solids
  • Blocking Distance: 0.4m (1.3ft) for both liquids and solids
  • Certifications: International explosion protection certificates

 


 

Benefits

  • Reliable Performance: Ensures accurate non-contact measurement.
  • Easy Setup: Quick and straightforward commissioning with a menu-guided on-site operation, available in seven selectable languages.
  • Diagnostic Display: On-site display of envelope curves enables easy troubleshooting.
  • Durable Design: Hermetically sealed and potted sensor provides long-term reliability.
  • Chemical Resistance: Made from PVDF, the sensor is resistant to chemical exposure.
  • Efficient Calibration: Allows calibration without the need for filling or emptying the tank.
  • Automatic Temperature Compensation: Integrated temperature sensor adjusts for variations in sound velocity due to temperature changes.

SPECIFICATIONS

Pressure Specifications

Measuring principle Ultrasonic
Characteristic / Application Compact ultrasonic transmitter
Supply / Communication

2-wire HART

Accuracy +/- 4 mm or +/- 0,2 % of set measuring range
Ambient temperature

-40 °C ... 80 °C

(-40 °F ... 176 °F)

Process temperature

-40 °C ... 80 °C

(-40 °F ... 176 °F)

Process pressure / max. overpressure limit

0.7 bar ... 2.5 bar abs

(10 psi ... 36 psi)

Main wetted parts PVDF
Process connection

Flange

DN80, ASME 3", JIS 10K 80

DN100,ASME 4", JIS 10K 100

Mounting bracket

Blocking distance 0.4 m (1.3 ft)
Max. measurement distance

Liquids: 10 m (33 ft),

Solids: 5 m (16 ft)

Communication 4...20 mA HART
Certificates / Approvals ATEX, FM, CSA, INMETRO, NEPSI
Application limits

Foam / high turbulence possible:

FMU44/FDU92

Fast filling and discharging

rate:

FMU90 + FDU9x

Level limit detection

FMU90 + FDU9x

Continuous / Solids Specifications

Measuring principle

Ultrasonic
Characteristic / Application Compact ultrasonic transmitter
Supply / Communication 2-wire HART
Accuracy +/- 4 mm or +/- 0,2 % of set measuring range
Ambient temperature

-40 °C ... 80 °C

(-40 °F ... 176 °F)

Process temperature

-40 °C ... 80 °C

(-40 °F ... 176 °F)

Process pressure / max. overpressure limit

0.7 bar ... 2.5 bar abs

(10 psi ... 36 psi)

Main wetted parts PVDF
Process connection

Flange

DN80, ASME 3", JIS 10K 80

DN100,ASME 4", JIS 10K 100

Mounting bracket

Blocking distance 0.4 m (1.3 ft)
Max. measurement distance

Liquids: 10 m (33 ft),

Solids: 5 m (16 ft)

Communication 4...20 mA HART
Certificates / Approvals ATEX, FM, CSA, INMETRO, NEPSI
Application limits Take notice of range diagram

 