Endress+Hauser
Digital Combined pH/ORP Sensor Memosens CPS96E
- SKU:
- Memosens CPS96E Series
- MPN:
- Memosens CPS96E
- Weight:
- 0.00 Grams
Description
Application
The Memosens CPS96E is ideal for applications involving media with a high concentration of suspended solids, such as dispersions, precipitates, or emulsions. Its open aperture design minimizes fouling, ensuring accurate and reliable pH and ORP measurements simultaneously. With Memosens 2.0 digital technology, the CPS96E offers enhanced process reliability and ease of use. It is moisture-resistant, supports lab-based calibration, and provides advanced storage for calibration and process data, making it an excellent choice for predictive maintenance. Industries and Applications:
- Chemical Processes
- Pulp and Paper Industry
- Contaminated Media and Suspended Solids:
- Dispersions
- Precipitates
- Emulsions
The CPS96E is certified for use in hazardous areas (zones 0, 1, and 2) with approvals including ATEX, IECEx, CSA C/US, NEPSI, Japan, and INMETRO.
Benefits
Memosens 2.0 offers advanced data storage capabilities for calibration and process information, facilitating better trend analysis and providing a solid foundation for predictive maintenance and enhanced IIoT integration. The sensor allows simultaneous measurement of pH, ORP, and rH values (in rH mode), offering a comprehensive view of process conditions for improved control. The integrated platinum electrode ensures consistent monitoring of reference impedance, maintaining high sensor quality. Its robust design, featuring stabilized gel and an ion-trap reference, makes it highly resistant to contamination and poisoning. The sensor remains unaffected by pressure and temperature fluctuations, as the open aperture prevents the formation of diffusion potentials at the junction. The non-contact inductive signal transmission ensures maximum process safety, while reduced process downtimes and an extended sensor lifespan contribute to lower operating costs.
Documents
You may want to read this product's catalog. If so, please click to view the Memosens CPS96E Series datasheet.SPECIFICATIONS
pH Specification
Measuring principle | Potentiometric |
Application | pH / ORP sensor for heavily contaminated media and suspended solids: - Chemical processes - Pulp and paper industry - Flue gas cleaning - Contaminated media: • Solids • Emulsions • Precipitation reactions |
Characteristic | Gel pH/ORP electrode with open aperture and double gel reference with ion trap for chemical applications |
Measuring range | ORP: –1 500 to 1 500 mV pH: 0 to 14 |
Measuring principle | Pt-disc as additional ORP element rH measurement and control of the reference impedance |
Design | All shaft lengths with temperature sensor Advanced gel technology Hydrolysis stable gel |
Material | Sensor shaft: Glass to suit process pH membrane glass: Type B Metal lead: Ag/AgCl ORP measuring element: Platinum O-ring: FKM Process coupling: PPS fiber-glass reinforced Nameplate: Ceramic metal oxide |
Dimension | Diameter: 12 mm (0.46 inch) Shaft lengths: 120, 225, 360, and 425 mm (4.68, 8.77, 14.04, and 16.57 inch) |
Process temperature | 0 to 110 °C (32 to 230 °F) |
Process pressure | 0.8 to 14 bar (11.6 to 203 psi) absolute |
Temperature sensor | NTC 30k |
Ex certification | With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in hazardous areas Zone 0, Zone 1, and Zone 2 |
Connection | Inductive, digital connection head with Memosens 2.0 technology |
Ingress protection | IP68 |
ORP/Redox Specification
Measuring principle | Sensor ORP / Redox |
Application | pH / ORP sensor for heavily contaminated media and suspended solids: - Chemical processes - Pulp and paper industry - Flue gas cleaning - Contaminated media: • Solids • Emulsions • Precipitation reactions |
Characteristic | Gel pH/ORP electrode with open aperture and double gel reference with ion trap for chemical applications |
Measuring range | ORP: –1 500 to 1 500 mV pH: 0 to 14 |
Measuring principle | Pt-disc as additional ORP element rH measurement and control of the reference impedance |
Design | All shaft lengths with temperature sensor Advanced gel technology Hydrolysis stable gel |
Material | Sensor shaft: Glass to suit process pH membrane glass: Type B Metal lead: Ag/AgCl ORP measuring element: Platinum O-ring: FKM Process coupling: PPS fiber-glass reinforced Nameplate: Ceramic metal oxide |
Dimension | Diameter: 12 mm (0.46 inch) Shaft lengths: 120, 225, 360, and 425 mm (4.68, 8.77, 14.04, and 16.57 inch) |
Process temperature | 0 to 110 °C (32 to 230 °F) |
Process pressure | 0.8 to 14 bar (11.6 to 203 psi) absolute |
Temperature sensor | NTC 30k |
Ex certification | With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in hazardous areas Zone 0, Zone 1, and Zone 2 |
Connection | Inductive, digital connection head with Memosens 2.0 technology |
Ingress protection | IP68 |