null
Endress+Hauser

Highly Robust Ultrasonic Flowmeter Proline Prosonic Flow G 300

(No reviews yet) Write a Review
SKU:
Proline Prosonic Flow G 300 series
Weight:
0.00 Grams
Call us: +971 4 259 5133

Description

Application

The Prosonic Flow G is ideal for a wide range of gas applications, providing reliable flow measurement even with wet gas and varying gas properties and compositions. It features a pressure-rated sensor housing equipped with a rupture disc for enhanced safety. The compact transmitter offers high flexibility for operation and system integration, including easy access from one side, a remote display, and improved connectivity options. Heartbeat Technology ensures continuous compliance and process safety.

The ultrasonic measurement principle remains unaffected by changes in gas composition.

This flowmeter delivers accurate measurement of natural and process gases in industries like chemical and oil and gas.

Device Features:

  • Direct measurement of flow, pressure, and temperature
  • Wetted parts made of titanium or 316L stainless steel
  • Maximum measuring accuracy of 0.5%
  • Compact dual-compartment housing with up to three I/Os
  • Backlit display with touch control and WLAN access
  • Optional remote display

Benefits

  • Flexible device with customizable gas mixtures for demanding measurement applications
  • Maximum reliability even with humid or wet gases due to a sensor design that resists condensate
  • High-performance process control with real-time pressure and temperature compensation
  • Efficient multivariable solution with no pressure loss
  • Full access to process and diagnostic information via numerous, configurable I/Os
  • Simplified setup with freely configurable I/O functionality
  • Integrated verification through Heartbeat Technology for reliable monitoring

 

 

SPECIFICATIONS

Gas Specifications

 

Measuring principle

 

Ultrasonic flow

 

Product headline

 

Highly robust gas specialist for fluctuating process conditions with compact, easily accessible transmitter.

Flexible device with user-definable gas mixtures for demanding measuring tasks.

Accurate measurement of natural and process gas in the chemical as well as oil and gas industries.

 

Sensor features

 

Maximum reliability even with humid or wet gas – sensor design insensitive to condensate. High-performance process control – real-time pressure- and temperature-compensated values. Efficient solution – multivariable, no pressure loss.

Direct measurement: flow, pressure & temperature. Wetted parts: titanium / 316L. Maximum measuring accuracy: 0.5 %.

 

Transmitter features

 

Full access to process and diagnostic information – numerous, freely combinable I/Os. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.

Compact dual-compartment housing with up to 3 I/Os. Backlit display with touch control and WLAN access. Remote display available.

 

Nominal diameter range

 

DN 25 to 300 (1 to 12")

Wetted materials

 

Measuring tube: 1.4408/1.4409 (CF3M)

Transducer: 1.4404 (316, 316L), Titan Grade 2

 

Measured variables

 

Volume flow, corrected volume flow, mass flow, flow velocity, speed of sound, pressure, temperature, density, dynamic viscosity, energy flow, Wobbe index, methane fraction, calorific value, molar mass

Max. measurement error

 

Volume flow (standard):

‐ ±1.0 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)

‐ ±2 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)

 

Volume flow (optional calibration):

 

‐ ±0.5 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)

‐ ±1.0 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)

Corrected volume flow (standard):

‐ ±1.5 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)

‐ ±2.5 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)

Corrected volume flow (optional calibration):

‐ ±1.0 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)

‐ ±1.5 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)

Sound Velocity: ±0.2 % o.r.

 

Measuring range

 

Gas: 0.3 m/s to 40 m/s

 

Max. process pressure

 

0.7 to 101 bar a (10.15 to 1464.88 psi a)

 

Medium temperature range

 

-50 to 150 °C (-58 to +302°F)

-50 to 100 °C (-58 to +212°F) with integrated pressure cell

 

Ambient temperature range

 

-40 to 60 °C(-40 to +140 °F)

Optional: -50 to 60 °C(-58 to +140 °F)

 

Sensor housing material

 

Stainless Steel, 1.4404(316/316L), 1.4408/1.4409 (CF3M)

 

Transmitter housing material

 

AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L

Polycarbonate

 

Degree of protection

 

Compact version: IP66/67, type 4X enclosure.

Optional: External WLAN antenna: IP67

 

Display/Operation

 

4-line backlit display with Touch Control (operation from outside)

Configuration via local display and operating tools possible

Remote display available

 

Outputs

 

3 outputs:

4-20 mA HART (active/passive)

4-20 mA (active/passive)

Pulse/frequency/switch output (active/passive)

Double pulse output (active/passive)

Relay output

 

Inputs

 

Status input

4-20 mA input

 

Digital communication

 

HART, Modbus RS485

 

Power supply

 

24V DC

100 to 230 V AC

AC 100 to 230 V / DC 24 V (non harzardous area)

 

Hazardous area approvals

 

ATEX, IECEx, cCSAus, JPN, EAC, UK Ex, KC

Product safety

 

CE, C-tick
Functional safety Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511

Metrological approvals and certificates

 

Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)

Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a

 

Pressure approvals and certificates

 

PED, CRN

 

Material certificates

 

3.1 material

NACE MR0175/MR0103

 

 

 

Documents

You may want to read this product's catalog. If so, please click to view the Endress+Hauser Proline Prosonic Flow G 300 Series datasheet.
View AllClose