null
Endress+Hauser

iTHERM ModuLine TM121

(No reviews yet) Write a Review
SKU:
iTHERM ModuLine TM121 series
Weight:
0.00 Grams
Call us: +971 4 259 5133

Description

Application

The iTHERM ModuLine TM121 is an RTD or TC assembly that features fundamental sensor technology, designed for use in non-hazardous environments. It includes a standard extension neck and offers an economical solution. The devices can be optionally equipped with Bluetooth®-enabled head transmitters, providing 4 to 20 mA output or HART® communication, which enhances measurement accuracy and reliability compared to directly wired sensors.

  • Suitable for a wide range of applications
  • Ideal for use in non-hazardous areas
  • Measuring range: –50 to +650 °C (–58 to +1202 °F)
  • Pressure range: up to 50 bar (725 psi)
  • Protection rating: up to IP 68

 


Benefits

  • Cost-effective and dependable measurement
  • Simplified selection, installation, and maintenance process
  • Extensive variety of process connections available
  • Convenient and secure device setup and maintenance with Bluetooth® communication

 

SPECIFICATIONS

Pressure Specifications

Measuring principle Resistance Temperature Detector
Characteristic / Application

metric style

modular temperature assembly

universal range of application

with neck

incl. thermowell / protection tube (metal)

Thermowell / protection tube welded protection tube

Insert / probe

Basis Sensortechnologie
Outer diameter protection tube / Insert

Thermowell:

9x1,25 mm

11x2 mm

Max. immersion length on request

up to 1.000,0 mm (39'')
Material protection tube/ thermowell

316L (1.4404)

Process connection

Thread:

G1/2"

NPT1/2"

M20x1.5, M27x2, M33x2

Compression fitting:

NPT1/2", G1/2", G1"

Flansche:

DN25 DN40 B1

ASME 1" 150 RF

ASME 1" 300 RF

ASME 1 1/2" 150 RF

ASME 2" 150 RF

ASME 2" 300 RF

Tip shape straight
Operating temperature range

Basic PT100 TF:

-50 °C …200 °C

(-58 °F …392 °F)

Typ K:

max. 1.100 °C

(max. 2.012 °F)

Max. process pressure (static) depending on the configuration
Accuracy

Class A acc. to IEC 60751

Class B acc. to IEC 60751

Class standard acc. to ASTM E230

Class 2 acc. to IEC 60584-2

Response time

t90 around 100 s

depending on configuration

Integration head transmitter yes (4 … 20 mA; HART)

 

 

Documents

You may want to read this product's catalog. If so, please click to view the Endress+Hauser iTHERM ModuLine TM121 Series datasheet.
View AllClose