Endress+Hauser
Manual or Automatic Retractable Assembly Cleanfit CPA472D
- SKU:
- Cleanfit CPA472D Series
- MPN:
- Cleanfit CPA472D
- Weight:
- 0.00 Grams
Description
Application The Cleanfit CPA472D assembly is designed to withstand demanding process conditions. It offers a wide range of durable materials, ensuring safe sensor operation even in environments with high pressure, high temperatures, or aggressive media. This assembly enables sensor replacement without interrupting the process and supports automatic cleaning and calibration through an optional control system. Enjoy simplified, time-saving sensor management and extended sensor lifespan. The Cleanfit CPA472D is suitable for installation in tanks or pipes within the following industries:
- Chemical industry
- Power plants
- Biodiesel production
It can handle pressures up to 10 bar.
Benefits
- Exceptionally durable: Operates under pressures of up to 10 bar and temperatures reaching 140°C, including PEEK, PVDF, and conductive PVDF versions.
- Superior chemical resistance for extended operational life.
- Eliminates the need for process interruptions during electrode cleaning and calibration, reducing maintenance time and costs.
- Versatile adaptability to various applications with a wide selection of materials and process connections.
Documents
You may want to read this product's catalog. If so, please click to view the Cleanfit CPA472D Series datasheet.SPECIFICATIONS
pH Specification
Measuring principle | Potentiometric |
Application | Chemical industry, biodiesel plants, power plants, plant design, tanks, process vats or pipes |
Installation | Retractable holder |
Characteristic | For tanks and process vats and pipelines or pipes (min. DN80) The robust retractable assembly permits replacement of the electrode while the tank is full or under process conditions with pressures of up to 6 bar |
Design | Manual or pneumatical |
Material | Electrode holder, rinse chamber and raised face: PEEK, PVDF, conductive PVDF, Alloy C22, stainless steel 1.4571 Seals: EPDM, FPM (Viton) or FFKM (Kalrez®) |
Dimension | Immersion depth: 146 or 280 mm (5.75 or 7.87 inch) |
Process temperature | 0 to 140 °C (32 to 284 °F) |
Process pressure | 0 to max. 4 bar (0 to max. 58 psi) overpressure for manual actuation 0 to 10 bar (0 to 145 psi) overpressure for pneumatic actuation |
Connection | Process connections: Flange DN 50, DN 80, ANSI 2", JIS IS10K50A, internal thread G1¼ |
ORP/Redox Specification
Measuring principle | Sensor ORP / Redox |
Application | Chemical industry, biodiesel plants, power plants, plant design, tanks, process vats or pipes |
Installation | Retractable holder |
Characteristic | For tanks and process vats and pipelines or pipes (min. DN80) The robust retractable assembly permits replacement of the electrode while the tank is full or under process conditions with pressures of up to 6 bar |
Design | Manual or pneumatical |
Material | Electrode holder, rinse chamber and raised face: PEEK, PVDF, conductive PVDF, Alloy C22, stainless steel 1.4571 Seals: EPDM, FPM (Viton) or FFKM (Kalrez®) |
Dimension | Immersion depth: 146 or 280 mm (5.75 or 7.87 inch) |
Process temperature | 0 to 140 °C (32 to 284 °F) |
Process pressure | 0 to max. 4 bar (0 to max. 58 psi) overpressure for manual actuation 0 to 10 bar (0 to 145 psi) overpressure for pneumatic actuation |
Connection | Process connections: Flange DN 50, DN 80, ANSI 2", JIS IS10K50A, internal thread G1¼ |