MPM4736 Digital Pressure Transmitter: Precision Monitoring for Diverse Applications

The MPM4736 digital pressure transmitter is designed for efficient and cost-effective operation within application systems, providing precise measurement capabilities for both pressure and temperature. This versatile instrument is adept at monitoring pressure or liquid levels while simultaneously capturing temperature information from both the medium and the environment.

Key Features:

  1. High-Precision Multi-Parameter Measurement:

    • Employs a high-precision digital processing circuit and specialized algorithms to ensure accurate measurement.
  2. Extremely Low Power Consumption:

    • Operates in auto-sleep mode with a consumption of no more than 10uA, resulting in longer battery life and reduced maintenance costs.
  3. Wide Voltage Support:

    • Accommodates multiple power supply methods, enhancing flexibility in various operational scenarios.
  4. Standard Protocol for Convenient Networking:

    • Supports standard protocols, facilitating seamless integration into networked environments.

Specifications:

  • Measured Range: -1bar…0bar~0.1bar…100bar
  • Overpressure: ≤1.5 times FS or 1100bar (min. value is valid)
  • Total Accuracy: ±0.25%FS(-10℃~70℃)
  • Temperature Accuracy:
    • ±0.5℃(-20℃~80℃)
    • ±0.75℃(-30℃~-20℃)
  • Long-term Stability: ±0.25%FS/year
  • Compensation Temperature: -10℃~70℃
  • Operation Temperature: -30℃~80℃

The MPM4736 stands out as an energy-efficient solution, combining precision in measurement with extended battery life. Its adaptability to diverse industries such as environmental protection, papermaking, and hydraulic control systems makes it a reliable choice for applications requiring simultaneous temperature and pressure monitoring.

View AllClose

Additional Information

Measured range:
-1bar…0bar~0.1bar…100bar
Overpressure:
≤1.5 times FS or 1100bar (min. value is valid)
Total accuracy:
±0.25%FS(-10℃~70℃)
Temperature accuracy:
±0.5℃(-20℃~80℃)±0.75℃(-30℃~-20℃)
Long-term stability:
±0.25%FS/year
Compensation temperature:
-10℃~70℃
Operation temperature:
-30℃~80℃