Description

Application

The compact design of the Promass A makes it perfect for continuous process control in challenging environments. Its innovative self-drainable, single-tube system ensures precise measurement of both liquids and gases, even at the lowest flow rates and high pressures. Promass A 200 is equipped with cost-effective, genuine loop-powered technology and features an intrinsically safe design (Ex ia), providing maximum operational safety in hazardous areas. Additionally, its Heartbeat Technology ensures reliable and safe processes.

The measurement principle functions independently of the physical properties of the fluid, such as viscosity or density.

It is particularly suitable for applications involving minimal flow quantities, especially within the chemical industry.

Device Features:

  • Nominal diameter: DN 1 to 4 (1/24" to 1/8")
  • Process pressure: Up to 430.9 bar (6250 psi)
  • Medium temperature: Up to +205°C (+401°F)
  • Loop-powered technology
  • Durable dual-compartment housing
  • Safety certifications: Global approvals (SIL, hazardous area)

Benefits

  • Space-saving design – Compact and lightweight sensor for easy installation
  • Top-tier product quality – Self-drainable measuring tube design across all sizes
  • Enhanced process safety – Resistant to corrosive environments and internal blockages
  • User-friendly wiring – Separate connection compartment for easier setup
  • Safe operation – Display with touch control and background lighting eliminates the need to open the device
  • Built-in verification – Integrated Heartbeat Technology for continuous monitoring

 

SPECIFICATIONS

Pressure Specifications

Measuring principle Coriolis
Product headline

Genuine loop-powered flowmeter for accurate measurement of lowest flow rates.

Suitable for applications with smallest flow quantities in the chemical industry.

Sensor features

Space-saving installation – compact, lightweight sensor. Highest product quality – self-drainable measuring tube design in all line sizes. Optimum process safety – resistant to corrosive ambient conditions and internal clogging.

Nominal diameter: DN 1 to 4 (¹⁄₂₄ to ¹⁄₈"). Process pressure up to 430.9 bar (6250 psi). Medium temperature up to +205 °C (+401 °F).

Transmitter features

Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology.

Loop-powered technology. Robust dual-compartment housing. Plant safety: worldwide approvals (SIL, Haz. area).

Nominal diameter range DN 1 to 4 (¹⁄₂₄ to ¹⁄₈")
Wetted materials Measuring tube: stainless steel, 1.4435 (316/316L); Alloy C22
Measured variables Mass flow, density, temperature, volume flow, corrected volume flow, reference density
Max. measurement error

Mass flow (liquid): ±0.1 %

Volume flow (liquid): ±0.1 %

Mass flow (gas): ±0.35 % Density (liquid): ±0.0005 g/cm³

Measuring range

0 to 450 kg/h (0 to 16.54 lb/min)

Max. process pressure 430.9 bar (6250 psi)
Medium temperature range -50 to 205 °C (-58 to 401 °F)
Ambient temperature range -40 to 60 °C (-40 to +140 °F)
Sensor housing material Stainless steel, 1.4404 (316L)
Transmitter housing material

Stainless steel, CF-3M (316L, 1.4404)

AlSi10MG, coated

Degree of protection IP66/67, Type 4X enclosure
Display/Operation

4-line backlit display with touch control (operation from outside)

Configuration via local display and operating tools possible

Remote display available

Outputs

4‐20 mA HART (passive)

4‐20 mA (passive)

Pulse/frequency/switch output (passive)

Inputs None
Digital communication HART, PROFIBUS PA, FOUNDATION Fieldbus
Power supply

DC 18 to 35 V (4‐20 mA HART with/without

pulse/frequency/switch)

DC 18 to 30 V (20 mA HART, 4‐20 mA)

DC 9 to 32 V (PROFIBUS PA)

Hazardous area approvals ATEX, IECEx, cCSAus, INMETRO, NEPSI, UK Ex
Product safety CE, C-TICK
Functional safety Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
Metrological approvals and certificates

Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)

Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)

Pressure approvals and certificates CRN
Material certificates 3.1 material
Hygienic approvals and certificates 3-A, cGMP

Gas Specifications

 

Measuring principle Coriolis
Product headline

Genuine loop-powered flowmeter for accurate measurement of lowest flow rates.

Suitable for applications with smallest flow quantities in the chemical industry.

Sensor features

Space-saving installation – compact, lightweight sensor. Highest product quality – self-drainable measuring tube design in all line sizes. Optimum process safety – resistant to corrosive ambient conditions and internal clogging.

Nominal diameter: DN 1 to 4 (¹⁄₂₄ to ¹⁄₈"). Process pressure up to 430.9 bar (6250 psi). Medium temperature up to +205 °C (+401 °F).

Transmitter features

Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology.

Loop-powered technology. Robust dual-compartment housing. Plant safety: worldwide approvals (SIL, Haz. area).

Nominal diameter range DN 1 to 4 (¹⁄₂₄ to ¹⁄₈")
Wetted materials Measuring tube: stainless steel, 1.4435 (316/316L); Alloy C22
Measured variables Mass flow, density, temperature, volume flow, corrected volume flow, reference density
Max. measurement error

Mass flow (liquid): ±0.1 %

Volume flow (liquid): ±0.1 %

Mass flow (gas): ±0.35 % Density (liquid): ±0.0005 g/cm³

Measuring range 0 to 450 kg/h (0 to 16.54 lb/min)
Max. process pressure 430.9 bar (6250 psi)
Medium temperature range -50 to 205 °C (-58 to 401 °F)
Ambient temperature range -40 to 60 °C (-40 to +140 °F)
Sensor housing material Stainless steel, 1.4404 (316L)
Transmitter housing material

Stainless steel, CF-3M (316L, 1.4404)

AlSi10MG, coated

Degree of protection IP66/67, Type 4X enclosure
Display/Operation

4-line backlit display with touch control (operation from outside)

Configuration via local display and operating tools possible

Remote display available

Outputs

4‐20 mA HART (passive)

4‐20 mA (passive)

Pulse/frequency/switch output (passive)

Inputs None
Digital communication

HART, PROFIBUS PA, FOUNDATION Fieldbus

Power supply

DC 18 to 35 V (4‐20 mA HART with/without

pulse/frequency/switch)

DC 18 to 30 V (20 mA HART, 4‐20 mA)

DC 9 to 32 V (PROFIBUS PA)

Hazardous area approvals ATEX, IECEx, cCSAus, INMETRO, NEPSI, UK Ex
Product safety CE, C-TICK
Functional safety Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
Metrological approvals and certificates

Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)

Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)

Pressure approvals and certificates CRN
Material certificates 3.1 material
Hygienic approvals and certificates 3-A, cGMP

Density Concentration Specification

Measuring principle Coriolis
Product headline

Genuine loop-powered flowmeter for accurate measurement of lowest flow rates.

Suitable for applications with smallest flow quantities in the chemical industry.

Sensor features

Space-saving installation – compact, lightweight sensor. Highest product quality – self-drainable measuring tube design in all line sizes. Optimum process safety – resistant to corrosive ambient conditions and internal clogging.

Nominal diameter: DN 1 to 4 (¹⁄₂₄ to ¹⁄₈"). Process pressure up to 430.9 bar (6250 psi). Medium temperature up to +205 °C (+401 °F).

Transmitter features

Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology.

Loop-powered technology. Robust dual-compartment housing. Plant safety: worldwide approvals (SIL, Haz. area).

Nominal diameter range DN 1 to 4 (¹⁄₂₄ to ¹⁄₈")
Wetted materials Measuring tube: stainless steel, 1.4435 (316/316L); Alloy C22
Measured variables Mass flow, density, temperature, volume flow, corrected volume flow, reference density
Max. measurement error

Mass flow (liquid): ±0.1 %

Volume flow (liquid): ±0.1 %

Mass flow (gas): ±0.35 % Density (liquid): ±0.0005 g/cm³

Measuring range 0 to 450 kg/h (0 to 16.54 lb/min)
Max. process pressure 430.9 bar (6250 psi)
Medium temperature range -50 to 205 °C (-58 to 401 °F)
Ambient temperature range -40 to 60 °C (-40 to +140 °F)
Sensor housing material Stainless steel, 1.4404 (316L)
Transmitter housing material

Stainless steel, CF-3M (316L, 1.4404)

AlSi10MG, coated

Degree of protection IP66/67, Type 4X enclosure
Display/Operation

4-line backlit display with touch control (operation from outside)

Configuration via local display and operating tools possible

Remote display available

Outputs

4‐20 mA HART (passive)

4‐20 mA (passive)

Pulse/frequency/switch output (passive)

Inputs None
Digital communication HART, PROFIBUS PA, FOUNDATION Fieldbus
Power supply

DC 18 to 35 V (4‐20 mA HART with/without

pulse/frequency/switch)

DC 18 to 30 V (20 mA HART, 4‐20 mA)

DC 9 to 32 V (PROFIBUS PA)

Hazardous area approvals ATEX, IECEx, cCSAus, INMETRO, NEPSI, UK Ex
Product safety CE, C-TICK
Functional safety Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
Metrological approvals and certificates

Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)

Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)

Pressure approvals and certificates PED, CRN
Material certificates 3.1 material
Hygienic approvals and certificates 3-A, cGMP

 

Documents

You may want to read this product's catalog. If so, please click to view the Endress+Hauser Proline Promass A 200 Series datasheet.
View AllClose