null
Endress+Hauser

TEC420 Thermocouple thermometer

(No reviews yet) Write a Review
SKU:
TEC420 Thermocouple thermometer series
Weight:
0.00 Grams
Call us: +971 4 259 5133

Description

Application

The TEC420 Thermocouple Thermometer, known for its modularity and robust design, is versatile and suitable for a wide range of industries. The optional head-mounted transmitter supports multiple communication protocols, offering improved measurement accuracy and reliability over directly wired sensors. Its variety of process connections, dimensions, and materials ensures adaptability across diverse applications.

The TEC420's sheathed thermocouples can measure temperatures ranging from -40°C to 1100°C, depending on the thermocouple type. They are ideal for low-pressure applications where direct contact with the medium or the use of a thermowell is required. They are also suitable for scenarios demanding short immersion lengths. It is important to note that the connection head should not exceed a maximum operating temperature of 80°C when a transmitter is installed.

 


Benefits

  • Exceptional flexibility due to its modular design, featuring standard terminal heads compliant with DIN EN 50446 and customizable immersion lengths.
  • High insert compatibility and design conformity with DIN 43772 standards.
  • Simplified head transmitter selection with options including 4 to 20 mA analog output, HART®, PROFIBUS® PA, or FOUNDATION Fieldbus™ protocols.

 

SPECIFICATIONS

Pressure Specifications

Measuring principle Thermocouple
Characteristic / Application

metric style

modular temperature assembly

suitable for hazardous areas

process connection as compression fitting

without neck

Thermowell / protection tube

 without (not intended to use with thermowell)
 Insert / probe  mineral insulated (MI), flexible
 Outer diameter protection tube / Insert  6,0 mm (0,24'')
 Max. immersion length on request  up to 100.000,00 mm (3.937,01'')
 Material protection tube/ thermowell 1.4401 (316)

Alloy 600 (2.4816)

 Process connection  depending on configuration
 Operating temperature range  -40 °C ...1.100 °C

(-40 °F ...2.012 °F)

 Max. process pressure (static)  at 20 °C: 20 bar (290 psi)
 Accuracy  class 2 acc. to IEC 60584
 Response time  t50 = 3 s

t90 = 7 s

 Integration head transmitter  yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)
 Ex - approvals  ATEX
 Certification  SIL (transmitter only)

 

 

Documents

You may want to read this product's catalog. If so, please click to view the Endress+Hauser TEC420 Thermocouple thermometer Series datasheet.
View AllClose